Medical device classes: how the MDR classifies a device
In the previous article we looked at when an electronic product is a medical device. The next question is its class, because the class determines who assesses the product before you can sell it and how much work lies behind it. The Medical Device Regulation (MDR) has four classes and 22 rules for assigning them. For electronic equipment, knowing a few of them well is enough.
What the class decides
The MDR classifies devices into classes I, IIa, IIb and III according to their intended purpose and inherent risks, using the rules in its Annex VIII. The class sets how conformity is demonstrated:
- Class I. The manufacturer draws up the technical documentation and issues the EU declaration of conformity itself. There are three exceptions where a notified body is involved, but it only reviews one aspect of the device. If the device is supplied sterile, it reviews how sterile conditions are established and maintained. If it has a measuring function, it checks that the device meets the metrological requirements. If it is a reusable surgical instrument, it reviews everything related to reuse: cleaning, disinfection, sterilisation, maintenance, functional testing and the related instructions for use.
- Classes IIa and IIb. A notified body is involved. It assesses the quality management system and the technical documentation of at least one representative device per category (IIa) or per generic device group (IIb).
- Class III. A notified body carries out the most thorough assessment, either of the quality management system and technical documentation or through type examination and product verification.
In vitro diagnostic devices have their own regulation, (EU) 2017/746, with different classes (A, B, C and D). We do not cover them here.
How the rules are applied
Before the rules themselves, Annex VIII sets out some implementing rules worth keeping in mind:
- Classification is governed by the device's intended purpose: what the manufacturer says it does, in its instructions, labelling and promotional material.
- If several rules apply, the strictest one prevails.
- Accessories are classified separately from the device they are used with.
- Software that drives a device or influences its use has the same class as the device. Standalone software is classified in its own right.
Rules 1 to 4 cover non-invasive devices, 5 to 8 invasive devices, 9 to 13 active devices, and 14 to 22 are special rules. An active device is one that depends on a source of energy other than the human body or gravity, and acts by converting it. Almost any electronic medical device is active, so the rules that matter most are those for active devices.
The rules for active devices
The examples for each rule come from the classification guidance MDCG 2021-24, revised in April 2026. They are examples with a typical intended purpose: your device is classified by its own.
Rule 9: therapy with energy
Active therapeutic devices that administer or exchange energy are class IIa. They are class IIb if they may do so in a potentially hazardous way, given the nature, density and site of application of the energy. The guidance places muscle stimulators, TENS devices, external hearing aids and ultrasound equipment for physiotherapy in IIa. In IIb, it lists lung ventilators, baby incubators and electrosurgical generators.
Rule 10: diagnosis and monitoring
Active devices for diagnosis and monitoring are class IIa if they allow direct diagnosis or monitoring of vital physiological processes. They are class IIb if they monitor vital parameters whose variations could result in immediate danger to the patient, such as cardiac performance or respiration, or if they diagnose in clinical situations where the patient is in immediate danger.
The guidance makes this concrete with a useful distinction. Continuous surveillance of vital parameters in anaesthesia, intensive care or emergency care is IIb. Taking readings during routine check-ups or self-monitoring is IIa. In its examples, electrocardiographs, electronic thermometers and electronic blood pressure monitors are IIa, while apnoea monitors, including those for home care, and multi-parameter monitors with alarms in intensive care are IIb.
Rule 12: administering or removing substances
Active devices that administer medicinal products, body liquids or other substances to the body, or remove them, are class IIa, and class IIb if they do so in a potentially hazardous way. Suction pumps and feeding pumps are IIa in the guidance. Infusion pumps, ventilators and dialysis equipment are IIb.
Rule 13: all other active devices
All other active devices are class I. The guidance mentions electric wheelchairs, electric hospital beds and patient hoists.
Rule 22: when the device closes the loop
Active therapeutic devices with an integrated diagnostic function that significantly determines patient management are class III. The MDR gives closed-loop systems and automated external defibrillators as examples. The guidance adds closed-loop insulin delivery systems and infusion pumps with sensors that adapt the therapy. It is a rule to bear in mind from the architecture stage: adding a sensor that adjusts the therapy automatically can move a device from IIa or IIb to class III.
Deciding before you design:Design review and rescue
Software and firmware
A device's firmware is not classified separately: it takes the class of the device it drives. Rule 11 is for standalone software, such as an app that analyses data on its own:
- Software that provides information used to take diagnostic or therapeutic decisions is class IIa. It is class IIb if those decisions could cause a serious deterioration of health or a surgical intervention, and class III if they could cause death or an irreversible deterioration.
- Software that monitors physiological processes is class IIa, or class IIb for vital parameters with a risk of immediate danger.
- All other software is class I.
If your product combines a device and an app, it is worth asking what each part does and whether the app makes decisions on its own.
And in the United States
The FDA also has three classes, from I (lowest risk) to III (highest risk), but assigns them differently. It does not apply rules: each generic device type, around 1,700 of them across 16 medical specialties, has its class in the FDA's product classification database. Class I carries general controls, and most of it is exempt from 510(k) premarket notification. Class II adds special controls, and class III requires premarket approval (PMA).
That is why there is no direct equivalence: the FDA does not distinguish IIa from IIb, and a device is classified there by looking up its type, not by working through rules. If your product is going to both markets, do the two classifications separately. For a formal answer, the FDA offers the 513(g) request. We work for both the EU and the United States, and in practice the documentation they ask for is not that different: an existing certification makes the next one much easier.
How to use this before you design
- Write down the intended purpose in one sentence: what the device does, for whom and in what setting. It is what decides the class, and changing it late changes the class.
- Work through the rules for active devices and for software, and keep the strictest one that applies.
- Compare with the examples in the MDCG 2021-24 guidance.
- Document the justification. The technical documentation has to include the class and the rule or rules applied. If the notified body disagrees, the competent authority of the manufacturer's country decides.
In the lower classes there is some room for interpretation. In the higher ones there is none if you want to sell the product. We try to design every product as if it were a medical device, even when it will not be one: the extra documentation effort usually pays off in the long run. If you have doubts about your product's class, we can review it with you before it is designed.
Preparing for the assessment:Test and validation